Why an AC-Coupled Scope Tilts a Flat Pulse

Before connecting any input or output cables, connect the DHO804 rear Ground
Terminal to earth ground using its supplied ground cable. Its Type-C power
input does not earth the instrument through the power cord. Use the supplied
power adapter and keep the AFG1022 protective earth connected. Shared coaxial
shields provide the signal reference; they do not replace that earth lead.
Use only the stated low-voltage grounded source. See the article for the full
fixture and wiring order.

Save the raw simultaneous
oscilloscope CSV and complete observations-blank.csv separately for each group.
Normalized analyzer CSV is mode,time_s,ch1_v,ch2_v, with increasing timestamps
within a mode. Voltages are in V, times in s. The four required mode tags are
slow_a (20 Hz CH1 DC), slow_b (20 Hz CH2 DC), fast_a (100 Hz CH1 DC), and
fast_b (100 Hz CH2 DC). Keep the physical channel numbers in the same columns.

Analyze a record with:
  python3 scope_coupling.py capture.csv

Run synthetic host checks with Python 3.8 or later:
  python3 -B -m unittest -v test_scope_coupling

Exit 0 reports the four qualified fits. Exit 1 means unresolved, including
FixtureError from wrong headers, invalid fields or modes, nonnumeric or
out-of-range samples, time/order or missing-mode errors, and failed fixture
or fit comparisons. Exit 2 reports file-access, text-decoding or CSV-syntax
errors; argument-parser errors also exit 2. The results table is printed only
after all four records pass. The model and worksheet do not include hardware
observations. SVG uses editable geometry and plot paths.