LDT0-028K film and two discharge loads
=====================================

Before connecting any input or output cables, connect the RIGOL DHO804’s rear Ground Terminal to earth ground with the supplied ground cable. Use the supplied power adapter: the Type-C power input does not earth the oscilloscope through its power cord. The probe’s G0 signal-return connection does not replace this separate chassis earth lead. “No external power” below refers only to the passive piezo fixture, not to the oscilloscope.

Use the fixture article for the mechanical stop and polarity instructions.
The 4.7 nF capacitor stays fitted. Only the 2.2 MΩ shunt link changes.
Record two press-and-hold traces for each load. The input captures must
cover −40 to +80 ms relative to the identified end of the press. The
supplied blank worksheet is a four-run template, not a set of readings.
Append the acquired points in increasing time order per run. Use signed
millivolts at the 10× probe tip and milliseconds. Use the raw scope view
to exclude clipping, rebound or ongoing movement before making the fit.

  python3 piezo_tail.py observations.csv
  python3 test_piezo_tail.py

Exit 0 means the observed fitted tails satisfy the fixture screens.
Exit 3 reports at least one unresolved screen; see its JSON reasons.
Exit 2 means invalid input or file trouble. Host tests use synthetic data.