A multimeter may report nearly the same resistance for a short piece of wire and for its own shorted test leads. The contacts and leads are part of that measurement. This experiment separates the wires that carry current from the wires that sense voltage. You will deliberately move a sense point and see why the boundary of a measurement matters.
The build uses a current-limited 5 V supply, two specified resistors and one ordinary digital multimeter, abbreviated DMM. You measure three voltages in sequence and calculate a resistance ratio. It demonstrates the principle of four-wire, or Kelvin, sensing. It is a manual teaching fixture, with a stated error budget; its results are not a calibration certificate or a promise of milliohm accuracy.
Parts, limits and a visible starting point
- One Vishay MRS25000C1000FCT00: MRS25 axial metal-film resistor, 100 ฮฉ, ยฑ1%, rated dissipation 0.6 W under the datasheet’s thermal conditions. This is R1, the current reference and current-limiting resistor.
- One Vishay MRS25000C1008FCT00: the same series, 1.00 ฮฉ, ยฑ1%. This is the initial device under test, or DUT. It gives a larger and easier voltage than a very short copper wire.
- A regulated 5.0 V bench supply with its current limit set to 100 mA, short insulated force wires and separate voltage-sense probes or clips. A solderless breadboard can hold R1, but avoid treating its unknown contacts as the DUT’s exact endpoints.
- A DMM used only in DC voltage mode while the supply is on. DC means direct current. Its voltage input resistance must be known and at least 10 Mฮฉ on the selected ranges. It needs a millivolt range or enough displayed resolution for roughly 5โ100 mV, and a range that can read the reference’s roughly 5 V.
- Optional isolated wire samples for later trials. Keep the test circuit separate from batteries, powered boards, household wiring and vehicle circuits. Only the specified bench source energizes it.
Use the meter manual to identify its voltage input socket, COM return socket, range, accuracy conditions and input resistance. A displayed final digit is resolution, not guaranteed accuracy. A meter with a coarse 0.01 V step cannot resolve the roughly 5 mV predicted for a 0.1 ฮฉ DUT here. Do not compensate for that by increasing current arbitrarily.
The reference is not a special four-terminal resistor. Its nominal value and tolerance are known, and the much smaller resistance of its short connections must stay inside the practical uncertainty allowance. You can improve the reference value only by measuring it with a suitable instrument under appropriate conditions. Typing 100.000 into a program does not improve a ยฑ1% resistor.

Force current through one path; sense at another pair of points
Disconnect the supply while assembling the loop. Connect supply positive to R1, R1 to the DUT, and the DUT to supply negative. Keep R1 in series for every powered trial. It limits current even when the DUT has very low resistance. There is no microcontroller, amplifier or analog-to-digital converter in this fixture.
| Name | Connection and meaning |
|---|---|
| A and B | Voltage test points at R1’s two terminals. Red probe at A, black at B measures Vref, the voltage across the reference. Do not move the black probe to supply negative for this reading. |
| F+ and Fโ | The force contacts that put current into and out of the DUT. Their outside wires carry the full loop current. |
| S+ and Sโ | The sense points on the DUT side of the force contacts. Red at S+, black at Sโ measures VDUT. The material between these two points defines the resistance being investigated. |
| One DMM | Move both voltage probes between the A/B pair and the S+/Sโ pair. These are sequential measurements, not two simultaneous channels. |
A force lead and a sense lead can meet at the DUT, but they should reach it independently. If both probes attach to the same remote breadboard row, the intervening contact and wire may still lie inside the sensed voltage. The name โfour wireโ is not enough: inspect where the two sensing paths actually land.
For the initial 1 ฮฉ resistor, place the force clips farther out on its leads and the sense probes closer to its body. Do not bridge its two terminals with a clip. For a wire, mark the two sense positions so you can return to the same boundary. Any lead portion or contact between S+ and Sโ belongs to the measured object; the arrangement does not remove every resistance everywhere.
The sense leads carry the meter’s very small input current instead of the loop’s tens of milliamps. Their voltage drop is consequently much smaller. โSmallโ is not โzeroโ: a finite meter input still loads the circuit. With a 10 Mฮฉ voltage input, its parallel loading of a 100 ฮฉ reference is approximately 100/10000000 = 10 parts per million. This is small compared with a 1% reference in this experiment. Input leakage, range switching, offsets and unstable contacts can matter more.
Obtain current from the reference voltage
Ohm’s law relates voltage V, current I and resistance R: V = IR. The same loop current passes through R1 and the DUT, ignoring the small loading of the voltmeter. Therefore I = Vref/Rref and RDUT = VDUT/I. Combining them gives RDUT = Rref ร VDUT/Vref. Use volts for both measured voltages, or use the same other unit for both. Do not mix 25 mV with 5 V as the bare numbers 25 and 5.
For an ideal 5 V source, 100 ฮฉ reference and 0.5 ฮฉ DUT, I = 5/100.5 = 0.049751 A, about 49.75 mA. Then Vref โ 4.9751 V and VDUT โ 0.024876 V, or 24.876 mV. The ratio returns 0.5 ฮฉ. These are predictions, not measured readings from a prototype.
For the initial 1 ฮฉ DUT, the ideal current is 5/101 โ 49.50 mA and its voltage is about 49.50 mV. A 0.1 ฮฉ DUT would produce only about 4.995 mV; a 2 ฮฉ DUT about 98.04 mV. This is why we start with a known 1 ฮฉ resistor before attempting a low-resistance wire.
The source need not be a precision current source: the measured reference voltage tells us the current at the reference-reading time. However, one meter measures the two locations at different times. If current changes between them, the ratio can be wrong even when each displayed voltage is individually correct.
With the ideal 0.5 ฮฉ example, R1 dissipates IยฒR โ 0.2475 W and the DUT about 0.00124 W. R1 can warm appreciably. The 0.6 W datasheet rating is conditional on thermal limits; it does not predict its surface temperature in your mounting. Its specified temperature coefficient is ยฑ50 parts per million per kelvin, abbreviated ppm/K. A hypothetical 30 K change at that coefficient corresponds to 0.15% resistance change. That is an illustration of possible error, not a measured temperature rise or a guaranteed drift correction.
A repeatable first trial
- With power disconnected, confirm R1’s value and continuity, and check the passive DUT. A broken reference invalidates I = Vref/Rref; the calculator cannot diagnose every wiring or component fault.
- Set the supply to 5.0 V and its limit to 100 mA. Verify polarity, R1 in series and no connected battery or live DUT. Put both meter leads in the voltage/COM sockets and select DC volts. Never put a current-input lead across the supply.
- Enable the source. Measure across A/B and record Vref,before. Expect roughly 5 V for the small passive DUT, rather than an exact required target. If the source enters current limit or readings behave unexpectedly, disconnect and inspect the loop.
- Move the voltage probes to S+/Sโ, choose a suitable voltage range, and record VDUT with its sign and unit. Do not move the force contacts during this three-reading sequence.
- Return both probes to A/B and record Vref,after. Repeat the complete sequence if reference readings change materially or the DUT reading will not settle. Allow for meter range-change settling.
- Use the mean of the two reference readings for the illustrative estimate. Compare the result for the 1 ฮฉ resistor with its tolerance and your meter error allowance. A reassuring rounded answer does not validate an inadequate error budget.
Bracketing the DUT reading with reference readings is a check for drift. It does not prove that the current stayed between those endpoint readings at every instant. For moving contacts or a fluctuating source, use simultaneous suitable instruments or a more controlled measurement method. This manual fixture is particularly unsuitable for a resistance that changes while you move the probes.

Move a sense point, not the DUT
Disconnect power and move one sense connection outward so it includes an additional force connection or wire segment. Keep the force loop otherwise unchanged. Re-energize and repeat the three readings. The new ratio should describe the newly included path. Its actual change depends on your contacts and wire; no minimum change is guaranteed.
As a calculated example, suppose a 1 ฮฉ DUT and an extra 0.2 ฮฉ path are in series with the 100 ฮฉ reference. The current is 5/101.2 โ 49.407 mA. Sensing only the DUT gives about 49.407 mV and a 1 ฮฉ result. Sensing the DUT plus that extra path gives about 59.289 mV and a 1.2 ฮฉ result. The current is the same in these two ideal sense placements. The difference is the boundary of the voltage measurement.
A contact can also change resistance when you disturb it, so a real trial may mix a boundary change with a physical contact change. Mark locations, repeat the original placement and record scatter. Do not report the diagram’s hypothetical 0.2 ฮฉ as the resistance of an unmeasured clip.
Use an error interval, with its assumptions visible
Take an illustrative set of readings: reference before 4.975 V, DUT 0.02487 V, reference after 4.970 V. With a nominal 100 ฮฉ reference, the mean-reference ratio is approximately 0.50015 ฮฉ. That number contains more displayed digits than its uncertainty warrants.
For illustration only, assign an absolute DUT-voltage error bound of ยฑ0.0002 V, a reference-voltage error bound of ยฑ0.005 V, and ยฑ1% reference-resistance tolerance covering its actual conditions. Replace those voltage bounds with your own meter’s range-dependent specification, including counts, offsets and required operating conditions. Different voltage ranges can have different gain errors; using the same meter does not guarantee cancellation.
If the reference-derived current bounds also cover the time of the DUT reading, a conservative calculation gives approximately 0.49043โ0.50998 ฮฉ. The program calls these conditional bounds. It has not measured stability, temperature, probe placement, input loading or the real meter errors. Unmodelled errors do not become harmless because a program prints a neat interval.
Follow the interval arithmetic
Let er and ed be absolute voltage-error bounds, and ฯ the reference’s fractional tolerance. Imin = [min(Vbefore,Vafter) โ er]/[Rref(1 + ฯ)]. Imax = [max(Vbefore,Vafter) + er]/[Rref(1 โ ฯ)]. For a passive nonnegative resistance, Rmin = max(0,VDUT โ ed)/Imax and Rmax = (VDUT + ed)/Imin. The lower current bound must be positive.
Complete optional calculator
This standard-library Python 3 program reads manually entered voltages; it does not connect to a meter. Run python3 kelvin_ratio.py 4.975 0.02487 4.970. All three values are in volts. The optional flags document the assumed reference, tolerance and absolute voltage-error bounds. Its default 2% reference-drift rejection is a chosen repeat-the-trial rule, not a statement about instrument accuracy.
"""Manual low-voltage resistance-ratio calculator, no instrument control.
Intervals assume that the two reference readings bound current during the DUT
reading. Bracketing samples do not establish that assumption by themselves.
"""
import argparse
import json
import math
def _number(name, value, low, high):
if isinstance(value, bool) or not isinstance(value, (int, float)):
raise ValueError(name + " must be a finite number")
if not math.isfinite(value) or not low <= value <= high:
raise ValueError(name + " is outside the calculator's supported range")
return float(value)
def estimate(v_before, v_dut, v_after, *, reference=100.0, tolerance=0.01,
dut_error=0.0002, reference_error=0.005, drift_limit=0.02):
"""Voltages/errors in V, reference in ohm, fractional tolerance/drift.
Error arguments are absolute bounds chosen from the actual meter manual,
including range, resolution, offset and operating conditions. Defaults are
illustrative, not the specification of any multimeter.
"""
vb = _number("reference before", v_before, 0, 10)
va = _number("reference after", v_after, 0, 10)
vd = _number("DUT voltage", v_dut, 0, 10)
rr = _number("reference resistance", reference, 0.001, 1e6)
tol = _number("reference tolerance", tolerance, 0, 0.2)
ed = _number("DUT voltage error", dut_error, 0, 1)
er = _number("reference voltage error", reference_error, 0, 1)
limit = _number("drift limit", drift_limit, 0, 0.2)
if min(vb, va) <= er:
raise ValueError("Reference lower bound is zero: check current path and repeat")
average = (vb + va) / 2
drift = abs(va - vb) / average
if drift > limit:
raise ValueError("Reference changed too much; do not combine these readings")
i_min = (min(vb, va) - er) / (rr * (1 + tol))
i_max = (max(vb, va) + er) / (rr * (1 - tol))
return {
"resistance_ohm": rr * vd / average,
"conditional_lower_ohm": max(0, vd - ed) / i_max,
"conditional_upper_ohm": (vd + ed) / i_min,
"reference_current_a": average / rr,
"bracket_drift_fraction": drift,
"assumption": "Current during DUT reading lies within reference-derived bounds; "
"reference is intact and its stated tolerance covers its actual temperature.",
"not_measured": "No meter, wire, contact or physical temperature was sampled by this program."
}
def main(argv=None):
parser = argparse.ArgumentParser(description=__doc__)
parser.add_argument("before", type=float, help="Reference voltage before, V")
parser.add_argument("dut", type=float, help="DUT voltage, V (not mV)")
parser.add_argument("after", type=float, help="Reference voltage after, V")
parser.add_argument("--reference", type=float, default=100)
parser.add_argument("--tolerance", type=float, default=0.01)
parser.add_argument("--dut-error", type=float, default=0.0002)
parser.add_argument("--reference-error", type=float, default=0.005)
parser.add_argument("--drift-limit", type=float, default=0.02)
a = parser.parse_args(argv)
try:
result = estimate(a.before, a.dut, a.after, reference=a.reference,
tolerance=a.tolerance, dut_error=a.dut_error,
reference_error=a.reference_error, drift_limit=a.drift_limit)
except ValueError as error:
parser.error(str(error))
print(json.dumps(result, indent=2, allow_nan=False))
if __name__ == "__main__":
main()
Download the complete calculator (ZIP; file: batch-005/code/kelvin_ratio.py) ยท Download the blank observation sheet (ZIP; file: batch-005/code/kelvin-wire-observations-blank.csv)
What failures should teach you
- Zero reference voltage: the loop may be open, the source off or the probes misplaced. Current cannot be inferred from a zero lower bound, so the calculator refuses the combination.
- Negative DUT voltage: check polarity, zero offset and the actual circuit. The program does not silently take its absolute value. Close to zero, meter offset or thermal voltages may dominate a tiny resistive drop.
- Zero DUT reading: this can mean โbelow the meter’s resolution.โ The lower bound can be zero while the upper bound remains positive. It does not establish a perfect conductor.
- Large reference drift: repeat the complete sequence and inspect the setup. The calculator cannot correct an unseen time-varying current.
- A very plausible but wrong result: inspect force/sense placement and verify the intact reference. A program checks its numbers; it does not see a probe on the wrong side of a clip.
Challenge: will adding resistance to a force lead always increase the calculated DUT resistance?
No. If the additional resistance lies outside the unchanged sense boundary, the loop current decreases, but both measured reference and DUT voltages decrease proportionally in the ideal stable case. Their ratio still gives the same DUT resistance. Moving the sense boundary to include that added path changes the answer. Sequential drift and real contact changes can spoil this ideal cancellation.
Record evidence before extending the fixture
Record the resistor markings and reference value used, meter model and both voltage ranges, error bounds and their source, supply settings, sense-point sketch, reference-before/DUT/reference-after sequence and repeated results. Note whether R1 warmed, whether the source remained in voltage regulation and whether touching a clip changed the reading. Keep calculated examples separate from measured rows.
For smaller resistance, the useful voltage gets smaller while offset, thermal electromotive voltages and meter uncertainty become more important. An electromotive voltage is a small voltage generated independently of the intended resistive current, for example at dissimilar-metal junctions at different temperatures. Dedicated instruments use further techniques, including controlled current and offset compensation. This experiment demonstrates why placement matters without claiming to reproduce every feature of those instruments.
Primary references and related reading
- Vishay MRS16/MRS25, document 28724, data revision 7 March 2016; current manufacturer-hosted PDF and part-number grammar
- Vishay MRS16/MRS25 current product documentation
- Keysight: separate force and sense connections, placement and stable test conditions
- Keithley/Tektronix application note 3176: two-wire and four-wire resistance measurement
Continue with measuring current with a resistor and voltmeter, why a meter or load changes a divider, and the interactive circuit lessons.